Ball Wave Palm-sized Gas Chromatograph

A gas chromatograph is an analyzer for measuring the species and concentration of a mixed gas of multiple species.

Ball Wave successfully developed a gas chromatograph with dimensions of 100 × 100 × 100 mm and a weight of around 1 kg for space applications through the JAXA Space Exploration Innovation Hub Center by utilizing ball SAW sensor technology.

Furthermore, they developed a portable palm-sized (A5 size) gas chromatograph called Sylph for terrestrial applications, and started providing prototypes. Thay also received “JAXA COSMODE” usage licensing from JAXA.

The basic principle used in Sylph is that the sample gas is directly sucked in by a pump, and is condensed by a built-in compact condenser. The carrier gas is provided by a hydrogen canister filled with a hydrogen absorbing alloy, and the mixed gas is injected into the column by a high-speed heater in the condenser. The various gas components separated by the column are detected by the ball SAW sensor based on the different retention times.

 

 

Category:
Description
Main body Model name SY-402
Size W133mm×H88mm×D174mm(Protrusions not included)
Weight 1.9kg
Power Requirements DC24V (port 5.5×2.1mm) ※AC adaptor100~240VAC、50/60Hz included)
Power Consumption Max. 72W
Connection USB
Environment 10℃~40℃、with no water condensation
Gas inlet Connector: 10-32 coned
Gas outlet Connector: 10-32 coned、Smell-sniffing nose patch connectable(Option)
Control PC Surface Pro 8、Windows 11
Operation Built-in control program
Start-up time Within 10 min after launching operation(including backflush)
Shut-down time Within 5 min after stop operation
Data output CSV file format
Carrier gas Type of gas Hydrogen1)
Flow rate 1ml / min
Supply method Hydrogen storage alloy canister2,3)
Capacity 6L
Estimated time of use 100 hours / canister
Concentrator Adsorbent Choice from TenaxRTA or Tenax TA+CarboxenR1000
Captured volume 1~1000ml、programmable
Injection temperature Up to 240℃
Cleaning Automatic at start-up and after measurement
Column Stationary layer Choice from High polarity, Mid polarity or Low polarity type4)
Temperature range 40℃~200℃
Hearing rate Max. 20℃ / min、programmable
Cooling rate Within 5 min from 200℃ to 50℃
Cleaning With backflush feature
Detector detector Ball SAW sensor
Sensing film Polydimethylsiloxane
Operating frequency 150MHz

 

Additional Information
Method

461-C-84-03, ASTM D-1160