KEM Karl Fischer Moisture Titrator (Coulometric titration) MKC-710B

The MKC‑710B is a compact and cost-effective coulometric Karl Fischer moisture titrator designed for precise trace-level water content determination in solids, liquids, and gases. It’s part of KEM’s 7th-generation MKC‑710 series, offering fast and accurate measurements with a compact footprint and a straightforward user interface.

Categories: , ,
Description
Type Karl Fischer Moisture Titrator
Model MKC-710B
Standard configuration MKC-710+IDP-100+Manual Solvent Change Unit
Measuring method Karl Fischer Coulometric titration
Measuring range Water content : 1ug to 300mg (depends on reagent)
Bromine content : 8ug to 300mg *1
Measurement cell 2-Component or 1-Component
Precision 0.1μg
Control method Constant current pulse time contro
Endpoint detection Alternate current polarization method with a twin platinum electrode
EP sense method Selective drift stability or limit measurement time
Required solvent 2-Component cell : Anolyte 100mL , Catholyte 5mL
1-Component cell : Anolyte 150mL
Method 20
Key operation Sheet key
Displays 1) Black and white LED-backlit LCD
2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish
3) 1-channel display
Calculation Concentration of water content , statistics data processing (mean , SD and RSD) and automatic averaging of blank value and factor value
Data storage 100 samples
GLP conformance Registration of operator /Record of check results / Record of factor measurement / Management of conduction time
External I/O RS-232C port x2 :
for Dot matrix printer , Electronic balance , Data Capture Software (SOFT-CAP)USB×1 :
for USB flash drive , Thermal printer , Keyboard , Barcode reader , Foot switch , USB HUB , Android device
Ambient condition 1) Temperature : 5 to 35°C
2) Humidity : 85%RH or below (no condensation)
Power source AC100 to 240V +/- 10% 50/60 Hz
Power consumption Main unit : Approx. 20W
Printer : Approx. 7W
Dimensions Titration unit : 141(W) x 292(D) x 244(H) mm
Stirrer : 107(W) x 206(D) x 340(H) mm (not incl. Solvent Change unit)
Solvent Change Unit : 240(W) x 140(D) x 405(H) mm (not incl. tubing)
Printer : 106(W) x 180(D) x 88(H) mm
Weight Titration unit : Approx.3.0kg
Stirrer : Approx.2.0kg
Solvent Change Unit : Aprox.0.6kg
Printer : Aprox.0.4kg
Conformity standard CE marking EMC : EN61326-1 LVD : EN61010-1
Additional Information
Make

KEM

Method

ASTM D1123, ASTM D1533, ASTM D3401, ASTM D4672, ASTM D4928, ASTM D5460, ASTM D6304, ASTM D6869, ASTM E1064, ISO 10101-3, ISO 12937, ISO 14897, ISO 15512, ISO 760, JIS C2101, JIS K0068, JIS K0113