The Karl Fischer Moisture Titrator MKC‑710M is a high-precision coulometric water-content analyzer from KEM. Designed for trace-level moisture determination in solids, liquids, and gases, it measures water content from 1 µg to 300 mg (up to 8 µg–300 mg for the bromine index) using Karl Fischer coulometric titration.
Touch‑screen control unit: 8.4″ color TFT LCD (800 × 600), supports multiple languages and simultaneous 4-channel display (including potentiometric or volumetric titrators).
High precision & speed: Resolution of 0.1 µg, RSD < 0.3% (n=10), and electrolytic speeds up to ~2.6 mg H₂O/min to reduce titration time.
Low power consumption: 30% reduction in energy use compared to prior models.
Dual‑cell flexibility: Supports both 1‑component (150 mL anolyte) and 2‑component (100 mL anolyte + 5 mL catholyte) cells.
Simultaneous multi‑unit operation: Connect up to four units (coulometric, volumetric, or potentiometric) via LAN or Bluetooth, with a single master control.
GLP compliance: User/group management, checks & reminders (performance, reagents, expiry dates), data storage for up to 500 samples, and statistical analysis (mean, SD, RSD).
Endpoint detection: Uses twin‑platinum electrode with alternate‑current polarization and drift correction or timed cutoff.
Integrated solvent handling: Comes with manual or optional automatic solvent change units, plus compatibility with evaporators (e.g., ADP‑611) and sample changers (CHK‑501).
Flexible connectivity: Ports include RS‑232 (×4), USB (for flash drives, printers, keyboards, barcodes), and LAN for PC communication.
Energy‑saving diaphragm: Replaceable ceramic diaphragm and optimized electrolytic control.
Type | Karl Fischer Moisture Titrator |
---|---|
Model | MKC-710M |
Standard configuration | MCU-710M+MKC-710+IDP-100+Manual Solvent Change unit |
Measuring method | Karl Fischer Coulometric titration |
Measuring range | Water content: 1ug to 300mg (depends on reagent) Bromine content: 8ug to 300mg *1 |
Measurement cell | 2-Component or 1-Component |
Precision | Relative standard deviation : less than 0.3% (n=10) *Per KEM standard measurement conditions and standard liquids |
Display resolution | 0.1ug |
Control method | Constant current pulse time control |
Endpoint detection | Alternate current polarization method with a twin platinum electrode |
EP sense method | Selective drift stability or limit measurement time |
Required solvent | 2-Component cell: Anolyte 100mL, Catholyte 5mL 1-Component cell: Anolyte 150mL |
Method | 120 |
Key operations | Touch panel |
Displays | 1) 8.4-inch color LCD 800 x 600 dots 2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish / German / French 3) Simultaneous 4-channel display (Can also display Automatic Potentiometric Titrator simultaneously) |
Calculation | Concentration of water content, statistical data processing (mean, SD and RSD) and automatic averaging of blank value and factor value |
Data storage | 500 samples |
GLP conformance | Registration of operator / User group administration Check performance with standard substance : Reminder of scheduled check date / Record of check results Reagent life control : Reminder of expiration/Reminder reagent replacement date Management of conduction time : Display of operating time |
External I/O | RS-232C port x4: for Dot matrix printer, Electronic balance, Data Capture Software (SOFT-CAP), Evaporator, Multiple sample changerUSB ×1: for USB flash drive, Thermal printer, A4 printer, Keyboard, Barcode reader, Foot switch, USB HUBLAN ×1: for Personal computer (PC) |
Extensibility | Measuring instrument: Automatic Potentiometric Titrator (AT-710), Karl Fischer Moisture Titrator (MKV-710/MKC-710/MKH-710) Three of these instruments can be addedEvaporator ADP-611Multiple sample changer: CHK-501 |
Ambient condition | 1) Temperature : 5 to 35°C 2) Humidity : 85%RH or below (no condensation) |
Power source | AC100 to 240V +/- 10% 50/60Hz |
Power consumption | Main unit: Approx. 30W Printer: Approx. 7W |
Dimensions | Touch-on panel controller : 225(W) x 190(D) x 42(H) mm Titration unit : 141(W) x 292(D) x 244(H) mm Stirrer : 107(W) x 206(D) x 340(H) mm (not incl. Solvent Change unit) Solvent Change Unit : 240(W) x 140(D) x 405(H) mm (not incl. tubing) Printer: 106(W) x 180(D) x 88(H) mm |
Weight | Touch-on panel controller: Approx.1.5kg Titration unit: Approx.3.0kg Stirrer: Approx.2.0kg Solvent Change Unit: Approx.0.6kg Printer: Approx.0.4kg |
Conformity standards | CE marking EMC: EN61326-1 LVD: EN61010-1 |
Make | KEM |
---|---|
Method | ASTM D1123, ASTM D1533, ASTM D3401, ASTM D4672, ASTM D4928, ASTM D5460, ASTM D6304, ASTM D6869, ASTM E1064, ISO 10101-3, ISO 12937, ISO 14897, ISO 15512, ISO 760 |